1700 X-ray room

An open table design, easy to install & maintain. Designed, manufactured & supported by vets for vets.

Category:

High-end materials 

Manufactured from high-quality anodized aluminum, The ClearRay 1700’s sturdy column can be adjusted to fit low ceilings

Functionality 

Unique 4-way floating table design with the greatest lateral range & the ability to accommodate up to 500 lbs. 

Open table design 

for convenience and easy cleaning 

Ease-of-use 

Foot controlled, one-touch function for  easier animal management and image capture. 

Table Specifications 

Fixed Arm Column for X-ray Tube and Collimator

Elevating table 18” to 31” from floor

Table Top 31”D x 60”L

Max Weight 500 lbs

Open Base Shape Table overall dimensions 60″x43.2″x 85″

Foot Switch for the Electromagnetic Brake System & collimator lights.

Standard Exposure Foot Switch (15 ft length)

 

NexGen Generator Specifications

————————————————————
Output Rating : 40 kW
————————————————————
Input Power : 230 VAC ±10%
50/60 Hz, Single-phase
————————————————————
kV Range : 40 – 125 kV (1 kV step)
————————————————————
mA Range : 10 – 500 mA
————————————————————
Time Range : 0.001 – 10 sec
————————————————————
Rotor Supply : Low Speed
————————————————————
Max. Power Output : 500 mA @ 80 kV
400 mA @ 100 kV
320 mA @ 125 kV
————————————————————
Reproducibility : Coefficient of Variation
kV < 0.005
Time < 0.005
mAs < 0.01
————————————————————
Accuracy : kV < ±(3% + 1 kV)
mA < ±(5% + 1 mA)
Time < ±(1% + 1 ms)
mAs < ±(5% + 1 mAs)
————————————————————
Linearity : Coefficient of Linearity < 0.01
CL = (X1 − X2) / (X1 + X2)
where X is mR/mAs
————————————————————

Clear Ray 1700 Brochure

ClearRay™-1700 Spec sheet

Pre-Install Sign Off For Clear Ray 1700 System