2000 Elevating X-ray room

The only 4-way elevating table on the market. Designed to facilitate x-rays for large patients.

Category:

High-end materials

​​Manufactured from high-quality anodized  aluminum, the sturdy column can be  adjusted to fit low ceilings 

Open table design 

for convenience and easy cleaning .

Ease-of-use 

Foot controlled, one-touch function for  easier animal management and image capture. 

Functionality 

Unique 4-way floating table design with  the greatest lateral range & the ability to  accommodate up to 500lbs. 

Electromagnetic braking system

Table Specifications 

Fixed Column ClearRay Stand for X-ray Tube and Collimator

Elevating table 18” to 31” from floor

Table Top 31”D x 60”L

Max Weight 500 lbs

Open Base Shape Table overall dimensions 60″x43.2″x 85″

Column Height 85”

Foot Switch for the Electromagnetic Brake System & collimator lights.

Foot Switch for exposure (15 ft length)

Foot Pedal for column control

 

NexGen Generator Specifications

————————————————————
Output Rating : 40 kW
————————————————————
Input Power : 230 VAC ±10%
50/60 Hz, Single-phase
————————————————————
kV Range : 40 – 125 kV (1 kV step)
————————————————————
mA Range : 10 – 500 mA
————————————————————
Time Range : 0.001 – 10 sec
————————————————————
Rotor Supply : Low Speed
————————————————————
Max. Power Output : 500 mA @ 80 kV
400 mA @ 100 kV
320 mA @ 125 kV
————————————————————
Reproducibility : Coefficient of Variation
kV < 0.005
Time < 0.005
mAs < 0.01
————————————————————
Accuracy : kV < ±(3% + 1 kV)
mA < ±(5% + 1 mA)
Time < ±(1% + 1 ms)
mAs < ±(5% + 1 mAs)
————————————————————
Linearity : Coefficient of Linearity < 0.01
CL = (X1 − X2) / (X1 + X2)
where X is mR/mAs
————————————————————

 

Collimator Specifications 

Collimator-CO-Vet Veterinarian Collimator

Electronic Timer

Automatic Laser Bucky Centering Light

CRV 2000 Brochure

ClearRay™-2000 Spec sheet

CRV 2000 Drawing Of Room Requirements

Pre-Install Sign Off For Clear Ray 2000 System